The new generation grating spectroradiometer MS-700N is designed to provide the most accurate solar spectral data outdoors. The MS-700N is a unique all weather sensor, without any moving parts. The detector core is temperature controlled to provide accurate irradiance measurement data within the spectral range from 350nm to 1050nm (Visible-NIR).
MS-700N is accurately calibrated with traceability to the International Standards and issued with a calibration uncertainty budget. The rugged optical design of the diffusor and input optics make the MS spectroradiometer concept superior to any fiber optic spectroradiometer which will be susceptible to mechanical vibration and handling. The spectroradiometers are designed for permanent installation, but are perfectly suited as a traveling reference.
- Reference for spectral VIS – NIR
- Optical resolution <10nm
- Operating temperature range -10 to 50°C
- Made for outdoor solar research
- Robust design no moving parts
Product Features
MS-700N has a separate power supply unit and can be controlled through RS232 / 422 by a PC or data logger. The PC software provides different functions for operating, data management and visualization. Through the open command protocol of the defined system control functions, software can be developed by the individual user. Measuring spectral irradiance is a must to understand the effect of the non-uniform energy distribution of the sun. Since the solar spectrum varies as a function of air-mass and composition of the atmosphere, the MS-700N reveals those details. While thermopile pyrheliometers and pyranometers are most suitable to quantify the total DNI or global radiation (W/m2), spectroradiometers give detail about the energy distribution (W/m2/nm), which is most important for PV or CPV cell research and performance analysis.
Main Features
1. All-Weather
MS-700N solar spectroradiometer is designed to be used as the spectral measurement standard in any outdoors environment around the world.
2. Durability
The dome, diffuser, grating and photodiode array are embedded in MS-700N to maximize the performance and guarantee durability.
3. Adaptive Measurement
The detector exposure time will be automatically adapted between 10ms and 5sec to optimize the signal to noise ratio, which makes the MS-700N suitable to measure during fluctuating solar irradiance conditions.
Technical Details
Wavelength range | 350 – 1050 nm |
Optical resolution FWHM | < 10 nm |
Wavelength accuracy | +/- 0.3 nm |
Directional response at 1000W/m² | < 7 % |
Temperature response -10°C to 50°C | < 2 % |
Temp. control | 25 °C |
Operating temperature range | -10 – 50 °C |
Exposure time | 10 – 5000 msec |
Dome material | Quartz |
Communication | RS-422 / 232C |
Power supply | 12VDC / 50VA |
Dimensions (mm) | 220 (D) x 197 (H) |
Weight | 4.5 kg |
Ingress protection IP | 65 |
Power supply (Power Adapter) | 100-240VAC, 50/60Hz |
Power consumption | 50 W |
Power supply operating conditions | -10 to 40°C / 0 to 90 %RH |
Power supply dimensions (mm) | 320 (W) x 240 (D) x 80 (H) |
Power supply weight | 1 kg |
Program | Analysis software WSDAc |
OS | Microsoft Windows 7/8/9/10 |
Functions | Display and analyze data |
Cable length | 10 m |
Field of View FOV | 180 ° |
Applications
- Measuring spectral irradiance
- Spectral irradiance portable solutions
- Measure spectral irradiance of sun simulator
- Spectral DNI Measurements